Abstract:
We have developed a class of aperture coding schemes for Remote Raman Spectrometers (RRS) that remove the traditional trade-off between throughput and spectral resolution. As a result, the size of the remote interrogation region can be driven by operational, rather than optical considerations. In this paper we present the design of our coded-aperture standoff spectroscopy system as well as experimental data collected while making remote measurements.