Publications [#268591] of James T. Dobbins

Papers Published
  1. Dobbins, JT; Rice, JJ; Beam, CA; Ravin, CE, Threshold perception performance with computed and screen-film radiography: implications for chest radiography., Radiology, vol. 183 no. 1 (April, 1992), pp. 179-187 [1549669], [doi] .

    Abstract:
    Images of a phantom obtained with computed radiography and standard screen-film imaging were compared to evaluate observer threshold perception performance with a modified contrast-detail technique. Optimum exposure necessary for performance with the imaging plate technique to match that with screen-film techniques was determined, as was comparative performance with variation in kilovoltages, plate type, spatial enhancement, and hard-copy interpolation method. It was found that computed radiography necessitates about 75%-100% more exposure than screen-film radiography to optimally match performance with Ortho-C film with Lanex regular or medium screens (Eastman Kodak, Rochester, NY) for detection of objects 0.05-2.0 cm in diameter. However, only minimal loss of detection performance (approximately 10% overall) was experienced if standard screen-film exposures were used with computed radiography. Little change in observer performance was found with variation in plate type, spatial enhancement, or method of hard-copy interpolation. However, perception performance with computed radiographic images was better at lower kilovoltages.