  1. Dobbins, J.T., III and Rice, J.J. and Beam, C.A. and Ravin, C.E., Threshold perception performance with computed and screen-film radiography: implications for chest radiography, Radiology (USA), vol. 183 no. 1 (1992), pp. 179 - 87 .

    Abstract:
    Images of a phantom obtained with computed radiography and standard screen-film imaging were compared to evaluate observer threshold perception performance with a modified contrast-detail technique. Optimum exposure necessary for performance with the imaging plate technique to match that with screen-film techniques was determined, as was comparative performance with variation in kilovoltages, plate type, spatial enhancement, and hard-copy interpolation method. It was found that computed radiography necessitates about 75%-100% more exposure than screen-film radiography to optimally match performance with Ortho-C film with Lanex regular or medium screens (Eastman Kodak, Rochester, NY) for detection of objects 0.05-20 cm in diameter. However, only minimal loss of detection performance (~10% overall) was experienced if standard screen-film exposures were used with computed radiography. Little change in observer performance was found with variation in plate type, spatial enhancement, or method of hard-copy interpolation. However, perception performance with computed radiographic images was better at lower kilovoltages