publications by Gregg E. Trahey.
Papers Published
- Trahey, G.E. and Hubbard, S.M. and von Ramm, O.T., Angle independent ultrasonic blood flow detection by frame-to-frame correlation of B-mode images,
Ultrasonics (UK), vol. 26 no. 5
,
pp. 271 - 6 [0041-624X(88)90016-9] .
(last updated on 2007/04/13)Abstract:
The authors have previously reported initial clinical results of a novel blood velocity imaging technique utilizing a two-dimensional correlation search applied to consecutively acquired echoes. In this paper, they describe both the physical principles underlying this technique and test tank experiments which define its performance under a variety of conditions. The results indicate that, unlike Doppler flow imaging systems, this technique defines the flow velocity vector in two dimensions and is not subject to aliasingKeywords:
acoustic imaging;acoustic signal processing;biomedical ultrasonics;flow measurement;haemodynamics;