publications by Robert Plonsey.
Papers Published
- Wen H. Ko and Bergmann, B.P. and Plonsey, R., Data acquisition system for body surface potential mapping,
J. Bioeng. (UK), vol. 2 no. 1-2
(1978),
pp. 33 - 46 .
(last updated on 2007/04/16)Abstract:
A system for acquiring digitally 180 simultaneous electrocardiographic surface potentials has been developed. Computer data processing yields up to 1000 equipotential contour maps per real-time second. Data is acquired from 180 active dry electrodes in an inflatable vest structure. The electronics include 12 amplification and sampling channels with associated digital control logic and a monitoring subsystem. Reproduced data channels are digitised synchronously by a PDP 11/45 A/D converter subsystem under external clock controlKeywords:
bioelectric potentials;computerised instrumentation;data acquisition;electrocardiography;voltage distribution;