publications by Robert Plonsey.


Papers Published

  1. Wen H. Ko and Bergmann, B.P. and Plonsey, R., Data acquisition system for body surface potential mapping, J. Bioeng. (UK), vol. 2 no. 1-2 (1978), pp. 33 - 46 .
    (last updated on 2007/04/16)

    Abstract:
    A system for acquiring digitally 180 simultaneous electrocardiographic surface potentials has been developed. Computer data processing yields up to 1000 equipotential contour maps per real-time second. Data is acquired from 180 active dry electrodes in an inflatable vest structure. The electronics include 12 amplification and sampling channels with associated digital control logic and a monitoring subsystem. Reproduced data channels are digitised synchronously by a PDP 11/45 A/D converter subsystem under external clock control

    Keywords:
    bioelectric potentials;computerised instrumentation;data acquisition;electrocardiography;voltage distribution;