| Publications [#133550] of Terry T Yoshizumi
Papers Published
- S Mukundan Jr, PI Wang, DP Frush, T Yoshizumi, J Marcus, E Kloeblen, M Moore, MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children.,
AJR. American journal of roentgenology, vol. 188 no. 6
(June, 2007),
pp. 1648-50, ISSN 1546-3141 [doi]
(last updated on 2013/07/01)
Abstract: OBJECTIVE: The purpose of our study was to measure radiation dose to the orbit during pediatric cranial CT with and without bismuth shielding using a novel dosimetry system. Cranial CT was performed on a pediatric anthropomorphic phantom, with and without bismuth eye shields. A solid-state metal oxide semiconductor field effect transistor (MOSFET) dosimeter was used to obtain real-time dose measurements. CONCLUSION: Bismuth shielding reduced radiation dose to the eye by up to 42%; shield artifact fell outside the diagnostic area of interest.
Keywords: Bismuth* • Body Burden • Child, Preschool • Equipment Design • Equipment Failure Analysis • Eye • Humans • Radiation Dosage • Radiation Protection • Radiometry • Relative Biological Effectiveness • Reproducibility of Results • Sensitivity and Specificity • Tomography, X-Ray Computed* • Transistors, Electronic* • instrumentation* • methods • methods* • radiography*
|