publications by G. Allan Johnson.
search www.ncbi.nlm.nih.gov.
Papers Published
- Fram, EK; Herfkens, RJ; Johnson, GA; Glover, GH; Karis, JP; Shimakawa, A; Perkins, TG; Pelc, NJ, Rapid calculation of T1 using variable flip angle gradient refocused imaging.,
Magn Reson Imaging, vol. 5 no. 3
(1987),
pp. 201-208 [3626789], [doi] .
(last updated on 2025/02/12)Abstract:
We present a method for rapid measurement of T1 relaxation times using gradient refocused images at limited flip angles and short repetition times. This "variable nutation" techniques was investigated using a T1 phantom. There was a high correlation between measurements obtained with the variable nutation and partial saturation techniques. The ability of this method to create calculated T1 images is also demonstrated. We conclude that the variable nutation method may allow measurement of T1 relaxation times with a significant reduction in acquisition time compared to partial saturation techniques.