publications by G. Allan Johnson.


search www.ncbi.nlm.nih.gov.

Papers Published

  1. Gewalt, SL; Glover, GH; Hedlund, LW; Cofer, GP; MacFall, JR; Johnson, GA, MR microscopy of the rat lung using projection reconstruction., Magn Reson Med, vol. 29 no. 1 (January, 1993), pp. 99-106 [8419748], [doi] .
    (last updated on 2025/02/14)

    Abstract:
    Projection reconstruction has been implemented with self-refocused selection pulses on a small bore, 2.0 T MR microscope, to allow imaging of lung parenchyma. Scan synchronous ventilation and cardiac gating have been integrated with the sequence to minimize motion artifacts. A systematic survey of the pulse sequence parameters has been undertaken in conjunction with the biological gating parameters to optimize resolution and signal-to-noise (SNR). The resulting projection images with effective echo time of < 300 microseconds allow definition of lung parenchyma with an SNR improvement of approximately 15 x over a more conventional 2DFT short echo gradient sequence.