publications by Patrick D. Wolf.
Papers Published
- Obeid, I; Nicolelis, MAL; Wolf, PD, A multichannel telemetry system for single unit neural recordings.,
Journal of Neuroscience Methods, vol. 133 no. 1-2
(February, 2004),
pp. 33-38 [14757342], [doi] .
(last updated on 2023/06/01)Abstract:
We present the design, testing, and evaluation of a 16 channel wearable telemetry system to facilitate multichannel single unit recordings from freely moving test subjects. Our design is comprised of (1) a 16-channel analog front end board to condition and sample signals derived from implanted neural electrodes, (2) a digital board for processing and buffering the digitized waveforms, and (3) an index-card sized 486 PC equipped with an IEEE 802.11b wireless ethernet card. Digitized data (up to 12 bits of resolution at 31.25k samples/s per channel) is transferred to the PC and sent to a nearby host computer on a wireless local area network. Up to 12 of the 16 channels were transmitted simultaneously for sustained periods at a range of 9 m. The device measures 5.1 cm x 8.1 cm x 12.4 cm, weighs 235 g, and is powered from rechargeable lithium ion batteries with a lifespan of 45 min at maximum transmission power. The device was successfully used to record signals from awake, chronically implanted macaque and owl monkeys.Keywords:
bioelectric potentials;biomedical electrodes;biomedical telemetry;cells (electric);local area networks;medical signal processing;microcomputers;neurophysiology;signal sampling;system buses;