publications by Wanda K. Neu.
Papers Published
- Dev, SB; Dhar, D; Krassowska, W, Electric field of a six-needle array electrode used in drug and DNA delivery in vivo: analytical versus numerical solution.,
Ieee Transactions on Bio Medical Engineering, vol. 50 no. 11
(November, 2003),
pp. 1296-1300 [TBME.2003.818467], [doi] .
(last updated on 2023/06/01)Abstract:
We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6%.Keywords:
biomedical electrodes;cancer;DNA;drug delivery systems;electric fields;finite element analysis;patient treatment;