| Publications [#63154] of Robert A Malkin
Papers Published
- Malkin, R.A. and Hoffmeister, B.K., AC leakage currents cause complete hemodynamic collapse below the ventricular fibrillation threshold,
Computers in Cardiology 1999. Vol.26 (Cat. No.99CH37004)
(1999),
pp. 351 - 3, Hannover, Germany [CIC.1999.825979]
(last updated on 2007/04/14)
Abstract: In 1993, the American standard for leakage current through the heart under a single-fault condition was increased from 10 μA to 50 μA rms. Both the original 10 μA standard and the new 50 μA standard were based on estimates of the ability of 60 Hz AC current to induce ventricular fibrillation (VF). This basis was a logical choice, since VF causes systemic blood pressure collapse, which can quickly cause morbidity and death. However, Swerdlow et al. (1999) showed that systemic pressure collapse can result from 60 Hz AC leakage currents (84±27 μA rms) far below those required to induce VF (278±226 μA rms) in 32 closed-chest humans. In this work, we extend the human results to a dog model and show preliminary evidence for the mechanism of the collapse
Keywords: ANSI standards;biomedical electrodes;electric shocks;electrocardiography;haemodynamics;leakage currents;safety;
|